MS Series Magnesense Differential Pressure Transmitter - Dwyer

The MS series Magnesense® differential pressure transmitters are versatile and can be used to monitor micro/differential pressure/vacuum, and to monitor air velocity/volume. Its functions include: metric/imperial engineering units can be selected on site, LCD digital display can be installed on site, the damping coefficient of the output signal can be adjusted, when combined with a pitot tube or other differential pressure wind speed sensor to measure the wind speed and air volume, the square root can be selected output.
Magnetic induction technology makes Magnesense® differential pressure transmitters have excellent performance, stability and reliability. Can be used for all differential pressure, air velocity and air volume measurement applications



Specification
item | value | Warranty | 1 year | Certification | IECEE, CCC, ISO | Customized support | OEM, ODM | Place of Origin | America | Brand Name | Dwyer | Model Number | MS-111 | Temperature Effect | 20 to 120°F (-6.67 to 48.9°C). | Output | 4-20MA | Range | High & Low | Mounting | Wall mounting | Application | Industries | |
Service: Air and non-combustible compatible gases.
Wetted Materials: Consult factory.
Accuracy: ±1% for 0.25" (50 Pa), 0.5" (100 Pa), 2" (500 Pa), 5" (1250 Pa), 10" (2 kPa), 15" (3 kPa), 25" (5 kPa) ±2% for 0.1" (25 Pa), 1" (250 Pa) and all bidirectional ranges.
Stability: ±1% F.S./year.
Temperature Limits: 0 to 150°F (-18 to 66°C).
Pressure Limits: 1 psi maximum, operation; 10 psi, burst.
Power Requirements: 10 to 35 VDC (2-wire); 17 to 36 VDC or isolated 21.6 to 33 VAC (3-wire).
Output Signals: 4 to 20 mA (2-wire); 0 to 5 V, 0 to 10 V (3-wire).
Response Time: Field adjustable 0.5 to 15 sec. time constant. Provides a 95% response time of 1.5 to 45 seconds.
Zero & Span Adjustments: Digital push button.
Loop Resistance: Current output: 0-1250 Ω max; Voltage output: min. load resistance 1 kΩ.
Current Consumption: 40 mA max.
Display (optional): 4 digit LCD.
Electrical Connections: 4-20 mA, 2-wire: European Style Terminal Block for 16 to 26 AWG. 0-10 V, 3-wire: European Style Terminal Block 16 to 22 AWG.
Electrical Entry: 1/2" NPS Thread. Accessory: Cable Gland for 5 to 10 mm diameter cable.
Process Connection: 3/16" (5 mm) ID tubing. Maximum OD 9 mm.
Enclosure Rating: NEMA 4X (IP66).
Mounting Orientation: Diaphragm in vertical position.
Weight: 8.0 oz (230 g).
Agency Approvals: CE.